DOI 10.1136/hrt.68.11.488

Медицинская библиотека · British heart journal · 1992

Transoesophageal echocardiography in neonates, infants and children: applicability and diagnostic value in everyday practice of a cardiothoracic unit.

Scott P J, Blackburn M E, Wharton G A, Wilson N, Dickinson D F, Gibbs J L

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To determine the applicability of and information obtained by transoesophageal echocardiography in neonates, infants, and children in every day practice of a cardiothoracic unit.

DESIGN

Four month prospective study.

SETTING

Supraregional centre for paediatric cardiothoracic services.

PATIENTS AND

METHODS

58 patients aged between four days and 16 years with a wide range of cardiovascular disease underwent transoesophageal echocardiography under sedation or general anaesthetic. One of two paediatric probes (6 or 7 mm diameter), or an adult probe (13 mm in diameter) was used. Whenever possible the investigation was immediately preceded by precordial echocardiography.

MAIN OUTCOME MEASURES: Success, failure, technical difficulties, and complications of probe introduction and, when possible, comparison of the information obtained with that obtained from precordial echocardiography.

RESULTS

Introduction of the probe was successful in 57 of the 58 patients. The only complication encountered was transient bradycardia during manipulation of the probe in a 2.2 kg baby. The adult (13 mm) probe was successfully used in children as small as 7.0 kg. Below this weight a smaller paediatric probe was required. In 56% of cases transoesophageal ultrasound provided information not obtained from the precordial approach. The technique was of particular value perioperatively and in the immediate postoperative period in neonates and infants and in the presence of valve prostheses and the investigation of mediastinal tumours in older children.

CONCLUSIONS

Transoesophageal echocardiography is a valuable additional investigative tool for children of all ages. It is of particular value when acquisition of precordial image is impaired around the time of and after operation and in children with prosthetic valves or mediastinal tumour. Image quality was superior with the adult probe and we recommend the use of this probe unless the patient's weight is below 7 kg, when a paediatric probe allows this technique to be used usefully and safely in babies as small as 2.2 kg.

Для цитирования: Scott P J, Blackburn M E, Wharton G A, Wilson N, Dickinson D F, Gibbs J L. Transoesophageal echocardiography in neonates, infants and children: applicability and diagnostic value in everyday practice of a cardiothoracic unit.. British heart journal. 1992. DOI: 10.1136/hrt.68.11.488.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.