DOI 10.1016/j.ijrobp.2023.12.036
LETd Optimization Verification With an SOI Microdosimeter.
Аннотация
PURPOSE: A first of its kind experimental verification of dose-averaged linear energy transfer (LETd) optimized treatment plans for proton therapy has been carried out using a silicon-on-insulator microdosimeter at the Massachusetts General Hospital (MGH), Boston, USA.
METHODS AND MATERIALS: Three clinical treatment plans of a typical ependymoma structure set were designed using the standard clinical approach, the proposed protocol approach, and a one-field approach. The plans were then reoptimized to reduce the LETd-weighted dose in the brain stem. All six plans were delivered in a solid water phantom and the experimental yD‾ measured.
RESULTS
After LETd optimization, a reduction in yD‾ was found within the brain stem by an average of 12%, 19%, and 4% for the clinical, protocol, and one-field plans, respectively, while maintaining adequate coverage of the tumor structure. The experimental LETd-weighted doses were in agreement with the treatment planning system calculations and Monte Carlo simulations and reinforced the improvement of the optimization.
CONCLUSIONS
This work demonstrates the first experimental verification of the clinical implementation of LETd optimization for patient treatment with proton therapy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.