Flow generators for helmet CPAP: Which to prefer? A bench study.

Privitera Daniele, Capsoni Nicolò, Zadek Francesco, Mazzone Annamaria, Bellone Andrea, Fumagalli Roberto, Langer Thomas

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To assess the different effect of filters' application during helmet-CPAP delivered with three different flow generators on the delivered fresh gas flow, FiO2, and the noise level inside and outside the helmet.

METHODS

In a bench study, three flow generators (air-oxygen blender, turbine ventilator and Venturi system) were used to generate two different gas flows (60 L/min and 80 L/min), with a fixed FiO2 at 0.6, to perform a helmet-CPAP on a manikin. Three different fixed PEEP valves (7.5, 10, and 12.5 cmH2O) were applied at the expiratory port. Gas flow, FiO2 and noise were recorded for each Flow-generator/Flow/PEEP combination, first without filter interposition and then after positioning a heat and moister exchanger filter (HMEF) at the helmet inlet port.

RESULTS

The application of the HMEF lead to a significant difference in the flow variation among the three flow generators (p < 0.001). Compared to baseline, the highest flow reduction was observed with the VENTURI (-13.4 ± 1.2 %, p < 0.001), a slight increase with the BLENDER (1.2 ± 0.5 %, p < 0.001), whereas no difference was recorded with the TURBINE (0.1 ± 0.6 %, p = 0.12). After HMEF was interposed, a significant FiO2 variation was observed only with VENTURI (11.3 ± 1.8 %, p < 0.001). As for the noise, the TURBINE was the least noisy system, both with and without the filter interposition.

CONCLUSIONS

Flow generators used to deliver helmet-CPAP have different characteristics and responses to HMEF interposition. Users should be aware of the effects on FiO2 and flow of the different devices in order to make a precise setup of the circuit.

Для цитирования: Privitera Daniele, Capsoni Nicolò, Zadek Francesco, Mazzone Annamaria, Bellone Andrea, Fumagalli Roberto, Langer Thomas. Flow generators for helmet CPAP: Which to prefer? A bench study.. Intensive & critical care nursing. 2022. DOI: 10.1016/j.iccn.2022.103344.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.