Respiratory impedance and bronchodilator responsiveness in healthy children aged 2-13 years.
Аннотация
BACKGROUND
The forced oscillation technique (FOT) can be used in children as young as 2 years of age and in those unable to perform routine spirometry. There is limited information on changes in FOT outcomes in healthy children beyond the preschool years and the level of bronchodilator responsiveness (BDR) in healthy children. We aimed to create reference ranges for respiratory impedance outcomes collated from multiple centers. Outcomes included respiratory system resistance (R(rs)) and reactance (X(rs)), resonant frequency (Fres), frequency dependence of R(rs) (Fdep), and the area under the reactance curve (AX). We also aimed to define the physiological effects of bronchodilators in a large population of healthy children using the FOT.
METHODS
Respiratory impedance was measured in 760 healthy children, aged 2-13 years, from Australia and Italy. Stepwise linear regression identified anthropometric predictors of transformed R(rs) and X(rs) at 6, 8, and 10 Hz, Fres, Fdep, and AX. Bronchodilator response (BDR) was assessed in 508 children after 200 µg of inhaled salbutamol.
RESULTS
Regression analysis showed that R(rs), X(rs), and AX outcomes were dependent on height and sex. The BDR cut-offs by absolute change in R(rs8), X(rs8), and AX were -2.74 hPa s L(-1), 1.93 hPa s L(-1), and -33 hPa s L(-1), respectively. These corresponded to relative and Z-score changes of -32%; -1.85 for R(rs8), 65%; 1.95 for X(rs8), and -82%; -2.04 for AX.
CONCLUSIONS
We have established generalizable reference ranges for respiratory impedance and defined cut-offs for a positive bronchodilator response using the FOT in healthy children.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.