Fink Christine, Jaeger Claudia, Jaeger Katharina, Haenssle Holger A

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

MelaFind is a multispectral computer vision system intended to -provide additional information on melanocytic lesions suspected of being melanoma by -objectively assessing their three-dimensional morphology.

OBJECTIVES

Analysis of the diagnostic performance of MelaFind in a real-life clinical setting.

PATIENTS AND

METHODS

In this observational study, 360 pigmented skin lesions (PSL) in 111 patients were assessed by office-based dermatologists using MelaFind. Scores ≥ 2 were considered to be suspicious of malignancy. The decision for surgical excision was left to the discretion of the examining dermatologists.

RESULTS

MelaFind scores ≥ 2 were observed in 147 of 360 PSL (40.8 %). Of the 107 excised lesions with a MelaFind-score ≥ 2, the diagnosis of melanoma was made in three cases; 53 (49.5 %) lesions proved to be dysplastic nevi. Among all lesions biopsied (n = 113), the sensitivity and specificity of MelaFind was 100 % and 5.5 %, respectively. While a higher specificity of 68.5 % may be assumed with respect to the overall data set (n = 360), this assumption is limited by incomplete follow-up data required to confirm that all non-excised lesions with a score < 2 were actually benign.

CONCLUSION

The high sensitivity of MelaFind facilitated the detection of melanoma. The overall specificity and benign-to-malignant ratio of excised lesions were acceptable. These parameters may be improved by using higher cutoff scores for excisional biopsies, and by more vigorously selecting PSL for MelaFind examination.

Для цитирования: Fink Christine, Jaeger Claudia, Jaeger Katharina, Haenssle Holger A. Diagnostic performance of the MelaFind device in a real-life clinical setting.. Journal der Deutschen Dermatologischen Gesellschaft = Journal of the German Society of Dermatology : JDDG. 2017. DOI: 10.1111/ddg.13220.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.