Experimental dosimetry in locally fabricated phantom for endobronchial brachytherapy plan: A phantom study.

Kant Ravi, Uniyal Satish Chandra, Rawat Jyoti, Gupta Meenu, Nautiyal Vipul

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study is to perform dosimetric analysis in a locally fabricated thorax phantom with the help of radiochromic film.

MATERIALS AND

METHODS

A thorax human body tissue equivalent phantom was fabricated. This phantom was scanned in the positron emission tomography/computed tomography (PETCT) machine and transferred the CT data into the Oncentra Master Plan treatment planning system (TPS). A brachytherapy (BT) plan created in TPS, exported to the high dose rate (HDR) BT machine. Placed Radiochromic EBT3 film dosimeter at the desired locations in the phantom and irradiate on machine with Ir-192 source. The TPS calculated values were compared with film measured values in the Phantom. Doses at five points in each film for each organ at risk were measured and compared with the TPS calculated doses to see the effects of distance on dose variation for analysis purpose.

RESULTS

The film measured doses deviate from TPS calculated doses with -2.5%, -3.2%, 3.8%, 2.4%, 5.3%, 17.75%, -7.1%, -9.4%, and 7.1% variation to Heart, Esophagus, Spinal Cord, descending Aorta, Ipsilateral Lung, Ipsilateral Lung 2 cm depth, pulmonary trunk, coronary artery, and contralateral Lung doses, respectively. In each film out of five points the center point and point closure to the source was having less variation between TPS calculated and film measured doses then other points.

CONCLUSION

This variation is due to the algorithm used in BT TPS so it is important to perform a verification of the treatment plan before execution in the patient to assure accuracy of treatment delivery. It is cheaper and used as a quality assurance Tool.

Для цитирования: Kant Ravi, Uniyal Satish Chandra, Rawat Jyoti, Gupta Meenu, Nautiyal Vipul. Experimental dosimetry in locally fabricated phantom for endobronchial brachytherapy plan: A phantom study.. Journal of cancer research and therapeutics. 2026. DOI: 10.4103/jcrt.jcrt_2095_25.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.