Are auditory steady-state responses a good tool prior to pediatric cochlear implantation?

Beck Roberto Miquelino de Oliveira, Grasel Signe Schuster, Ramos Henrique Faria, Almeida Edigar Rezende de, Tsuji Robinson Koji, Bento Ricardo Ferreira, Brito Rubens de

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

ASSR allow frequency-specific evaluation in intensities up to 120dB HL and detection of residual hearing in patients with severe-to-profound hearing loss.

AIM: to compare ASSR thresholds and behavioral test results in children with suspected severe-to-profound hearing loss.

METHODS

Cross sectional study to compare ASSR and behavioral responses (VRA or audiometry) in 63 pediatric cochlear implant candidates (126 ears) aged between 6 and 72 months. We included children with normal otomicroscopy, absent responses to click-ABR and otoaccoustic emissions. We excluded children with inner ear malformations, auditory neuropathy spectrum disorder or who did not complete VRA or achieve EEG noise<30nV during the ASSR test. Air-conduction ASSR stimuli were continuous sinusoidal tones presented at 0.5, 1, 2 and 4kHz starting at 110dB HL. Behavioral thresholds were acquired with warble tones presented at 0.5, 1, 2 and 4kHz in each ear through insert or head phones at maximum presentation level of 120dB HL.

RESULTS

Behavioral thresholds were obtained in 36.7% (185/504) of all frequencies in all subjects, 9% in intensities >110dB HL. Among 504 ASSR measurements, 53 thresholds were obtained (10.5%). Overall 89.5% of the tested frequencies did not show any response at 110dB HL. Most responses were at 500Hz. Mean differences between behavioral and ASSR thresholds varied from 0.09 to 8.94dB. Twenty-seven comparisons of behavioral and ASSR thresholds were obtained: 12 at 0.5kHz, 9 at 1kHz, 5 at 2kHz and 1 at 4kHz. Absent responses were observed in both tests in 38.1% at 0.5kHz, 52.4% at 1kHz, 74.6% at 2kHz and 81.0% at 4kHz. Specificity was>90% at 1, 2 and 4kHz. In ears with no behavioral response at 120dB HL all ASSR thresholds were in the profound hearing loss range, 90% of them were ≥110dB HL.

CONCLUSION

Among 63 pediatric CI candidates, absent responses to high-intensity ASSR was the major finding (specificity>90%) predicting behavioral thresholds in the profound hearing loss range. These findings can be helpful to confirm the decision for cochlear implantation.

Для цитирования: Beck Roberto Miquelino de Oliveira, Grasel Signe Schuster, Ramos Henrique Faria, Almeida Edigar Rezende de, Tsuji Robinson Koji, Bento Ricardo Ferreira, Brito Rubens de. Are auditory steady-state responses a good tool prior to pediatric cochlear implantation?. International journal of pediatric otorhinolaryngology. 2015. DOI: 10.1016/j.ijporl.2015.05.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.