Effects of radiation dose reduction in Volume Perfusion CT imaging of acute ischemic stroke.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Othman Ahmed E, Brockmann Carolin, Yang Zepa, Kim Changwon, Afat Saif, Pjontek Rastislav, Nikoubashman Omid, Brockmann Marc A, Kim Jong Hyo, Wiesmann Martin
Аннотация
PURPOSE: To examine the influence of radiation dose reduction on image quality and sensitivity of Volume Perfusion CT (VPCT) maps regarding the detection of ischemic brain lesions.
METHODS AND MATERIALS: VPCT data of 20 patients with suspected ischemic stroke acquired at 80 kV and 180 mAs were included. Using realistic reduced-dose simulation, low-dose VPCT datasets with 144 mAs, 108 mAs, 72 mAs and 36 mAs (80 %, 60 %, 40 % and 20 % of the original levels) were generated, resulting in a total of 100 datasets. Perfusion maps were created and signal-to-noise-ratio (SNR) measurements were performed. Qualitative analyses were conducted by two blinded readers, who also assessed the presence/absence of ischemic lesions and scored CBV and CBF maps using a modified ASPECTS-score.
RESULTS: SNR of all low-dose datasets were significantly lower than those of the original datasets (p < .05). All datasets down to 72 mAs (40 %) yielded sufficient image quality and high sensitivity with excellent inter-observer-agreements, whereas 36 mAs datasets (20 %) yielded poor image quality in 15 % of the cases with lower sensitivity and inter-observer-agreements.
CONCLUSION: Low-dose VPCT using decreased tube currents down to 72 mAs (40 % of original radiation dose) produces sufficient perfusion maps for the detection of ischemic brain lesions.
KEY POINTS: • Perfusion CT is highly accurate for the detection of ischemic brain lesions • Perfusion CT results in high radiation exposure, therefore low-dose protocols are required • Reduction of tube current down to 72 mAs produces sufficient perfusion maps.
METHODS AND MATERIALS: VPCT data of 20 patients with suspected ischemic stroke acquired at 80 kV and 180 mAs were included. Using realistic reduced-dose simulation, low-dose VPCT datasets with 144 mAs, 108 mAs, 72 mAs and 36 mAs (80 %, 60 %, 40 % and 20 % of the original levels) were generated, resulting in a total of 100 datasets. Perfusion maps were created and signal-to-noise-ratio (SNR) measurements were performed. Qualitative analyses were conducted by two blinded readers, who also assessed the presence/absence of ischemic lesions and scored CBV and CBF maps using a modified ASPECTS-score.
RESULTS: SNR of all low-dose datasets were significantly lower than those of the original datasets (p < .05). All datasets down to 72 mAs (40 %) yielded sufficient image quality and high sensitivity with excellent inter-observer-agreements, whereas 36 mAs datasets (20 %) yielded poor image quality in 15 % of the cases with lower sensitivity and inter-observer-agreements.
CONCLUSION: Low-dose VPCT using decreased tube currents down to 72 mAs (40 % of original radiation dose) produces sufficient perfusion maps for the detection of ischemic brain lesions.
KEY POINTS: • Perfusion CT is highly accurate for the detection of ischemic brain lesions • Perfusion CT results in high radiation exposure, therefore low-dose protocols are required • Reduction of tube current down to 72 mAs produces sufficient perfusion maps.