Evaluation of the Spot Vision Screener for children with limited access to ocular health care.

Barugel Raphaël, Touhami Sara, Samama Sacha, Landre Céline, Busquet Gauthier, Vera Liza, Bui Quoc Emmanuel

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the sensitivity, specificity, and referral rate of the Spot Vision Screener (Welch Allyn Inc, Skaneateles Falls, NY) with the gold standard cycloplegic measurements acquired using the Retinomax in a population of underprivileged children and teenagers with limited access to medical care.

METHODS

Children were recruited for the study by social workers in the vicinity of Robert Debre Hospital, Paris, France. Refractive errors (hyperopia of ≥ +2.00 D spherical equivalent [SE]; myopia of ≤ -0.50 D SE; astigmatism of ≥1.00 D between the two main meridians; anisometropia of ≥1.00 D SE difference between eyes) were assessed using the Spot Vision Screener and the Retinomax. Sensitivity (true positive rate), specificity (true negative rate), and referral rate of this Spot Vision screening program were evaluated.

RESULTS

A total of 82 eyes of 41 subjects (19 males) were included; mean age was 126 months of age (range, 48-246). The sensitivity of the Spot Vision Screener for the detection of refractive errors was 82.35%; specificity was 91.67%. The sensitivity of the Spot Vision Screener to detect hyperopia, myopia, astigmatism, and anisometropia was 27.27%, 84.61%, 78.57%, and 66.67%, respectively. Its specificity to detect hyperopia, myopia, astigmatism, and anisometropia was 100%, 98.55%, 89.71% and 94.29%, respectively.

CONCLUSIONS

The specificity of the Spot Vision Screener to detect refractive errors was found to be relatively high (>90%). However, its low sensitivity for hyperopia seems to remain a major limitation of the device, because hyperopia is particularly important to detect in children given its high prevalence and possible adverse consequences. Global programs using cycloplegic measurements should be considered an alternative.

Для цитирования: Barugel Raphaël, Touhami Sara, Samama Sacha, Landre Céline, Busquet Gauthier, Vera Liza, Bui Quoc Emmanuel. Evaluation of the Spot Vision Screener for children with limited access to ocular health care.. Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus. 2019. DOI: 10.1016/j.jaapos.2018.09.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.