DOI 10.1016/j.ijporl.2012.01.030
Intra- and postoperative electrically evoked stapedius reflex thresholds in children with cochlear implants.
Аннотация
OBJECTIVES
The aim of this study was to investigate whether there is a significant correlation between intra- and postoperative electrically evoked stapedius reflex thresholds (eSRTs) in children with cochlear implants.
METHODS
Sixty-five pediatric cochlear implant users were included in this study. All patients had congenital prelingual hearing loss. The round window approach was used in all patients. The eSRTs were intraoperatively measured using the 1st, 3rd, 6th and 12th electrodes of the cochlear implant. The measurements taken during the first fitting of the device were taken again one month after surgery. We used paired-sample t-tests to determine the correlation between intra- and postoperative eSRTs.
RESULTS
The eSRT analysis revealed a statistically significant difference between the intra- and postoperative thresholds. A correlation analysis did not reveal any correlation between intra- and postoperative eSRTs.
CONCLUSION
Intraoperative eSRT measurements were unable to predict early postoperative eSRTs.