Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2011
Evaluation of an electrode prototype for atraumatic cochlear implantation in hearing preservation candidates: preliminary results from a temporal bone study.
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Helbig Silke, Settevendemie Claudia, Mack Martin, Baumann Uwe, Helbig Matthias, Stöver Timo
Аннотация
OBJECTIVES: This study aimed to evaluate an atraumatic prototype electrode carrier for cochlear implantation, the FLEX electrode. This electrode is designed to preserve hearing and to achieve a 360-degree insertion.
STUDY DESIGN: A cross-sectional human temporal bone study was conducted.
SETTING: Preliminarily, the prototype electrode was inserted in a scala tympani model to measure the insertion force. Thirteen human temporal bones were acquired postmortem and implanted with the new device using the round window approach. Three of them were implanted under radiologic control to demonstrate the insertion path. After embedding, the remaining 10 temporal bones were sectioned undecalcified and examined macroscopically and histologically.
MAIN OUTCOME MEASURES: The insertion force was measured to determine intracochlear resistance peaks. The insertion angle was measured, and the degree of intracochlear trauma was determined.
RESULTS: The round window approach caused cochlear trauma in 1 of 10 specimens. An exact 360-degree insertion was achieved in 7 of 10 specimens (mean overall insertion angle, 360 degrees). Radiologic examination and insertion force measurements revealed the smooth and atraumatic insertion.
CONCLUSION: The new prototype electrode carrier is suitable for clinical application. It can be handled easily and allows atraumatic 360-degree insertion of all electrode contacts. Therefore, this electrode concept allows good coverage of the cochlea for electrical and additional acoustic stimulation.
STUDY DESIGN: A cross-sectional human temporal bone study was conducted.
SETTING: Preliminarily, the prototype electrode was inserted in a scala tympani model to measure the insertion force. Thirteen human temporal bones were acquired postmortem and implanted with the new device using the round window approach. Three of them were implanted under radiologic control to demonstrate the insertion path. After embedding, the remaining 10 temporal bones were sectioned undecalcified and examined macroscopically and histologically.
MAIN OUTCOME MEASURES: The insertion force was measured to determine intracochlear resistance peaks. The insertion angle was measured, and the degree of intracochlear trauma was determined.
RESULTS: The round window approach caused cochlear trauma in 1 of 10 specimens. An exact 360-degree insertion was achieved in 7 of 10 specimens (mean overall insertion angle, 360 degrees). Radiologic examination and insertion force measurements revealed the smooth and atraumatic insertion.
CONCLUSION: The new prototype electrode carrier is suitable for clinical application. It can be handled easily and allows atraumatic 360-degree insertion of all electrode contacts. Therefore, this electrode concept allows good coverage of the cochlea for electrical and additional acoustic stimulation.