Comprehensive analysis of cochlear implant failure: usefulness of clinical symptom-based algorithm combined with in situ integrity testing.

Yamazaki Hiroshi, O'Leary Stephen, Moran Michelle, Briggs Robert

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Accurate diagnosis of cochlear implant failures is important for management; however, appropriate strategies to assess possible device failures are not always clear. The purpose of this study is to understand correlation between causes of device failure and the presenting clinical symptoms as well as results of in situ integrity testing and to propose effective strategies for diagnosis of device failure.

STUDY

DESIGN

Retrospective case review.

SETTING

Cochlear implant center at a tertiary referral hospital.

PATIENTS: Twenty-seven cases with suspected device failure of Cochlear Nucleus systems (excluding CI512 failures) on the basis of deterioration in auditory perception from January 2000 to September 2012 in the Melbourne cochlear implant clinic.

MAIN OUTCOME MEASURES: Clinical presentations and types of abnormalities on in situ integrity testing were compared with modes of device failure detected by returned device analysis.

RESULTS

Sudden deterioration in auditory perception was always observed in cases with "critical damage": either fracture of the integrated circuit or most or all of the electrode wires. Subacute or gradually progressive deterioration in auditory perception was significantly associated with a more limited number of broken electrode wires. Cochlear implant mediated auditory and nonauditory symptoms were significantly associated with an insulation problem. An algorithm based on the time course of deterioration in auditory perception and cochlear implant-mediated auditory and nonauditory symptoms was developed on the basis of these retrospective analyses, to help predict the mode of device failure. In situ integrity testing, which included close monitoring of device function in routine programming sessions as well as repeating the manufacturer's integrity test battery, was sensitive enough to detect malfunction in all suspected device failures, and each mode of device failure showed a characteristic abnormality on in situ integrity testing.

CONCLUSION

Our clinical manifestation-based algorithm combined with in situ integrity testing may be useful for accurate diagnosis and appropriate management of device failure. Close monitoring of device function in routine programming sessions as well as repeating the manufacturer's integrity test battery is important if the initial in situ integrity testing is inconclusive because objective evidence of failure in the implanted device is essential to recommend explantation/reimplantation.

Для цитирования: Yamazaki Hiroshi, O'Leary Stephen, Moran Michelle, Briggs Robert. Comprehensive analysis of cochlear implant failure: usefulness of clinical symptom-based algorithm combined with in situ integrity testing.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2014. DOI: 10.1097/mao.0000000000000267.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.