STUDY

OBJECTIVE

Our objective was to evaluate a portable device (Somté, Compumedics, Australia), which incorporates 2 neurophysiological channels (electroencephalography and electrooculography) with cardiorespiratory monitoring for the diagnosis of obstructive sleep apnea (OSA).

METHOD: Full polysomnography (PSG) and Somté recordings were simultaneously performed in 68 patients with suspected OSA. Data were analyzed blindly by 2 scorers.

RESULTS

A good agreement between methods in sleep efficiency was observed (68.8% [18.4] with PSG vs 68% [19.1] with Somté [p: n.s.] for scorer 1, and 67.5% [19.1] vs 68.4% [18.5; p: n.s.] for scorer 2). The apnea-hypopnea index (AHI) obtained with Somté was lower than with PSG: 19 (17.8) vs 21.7 (19) (p < 0.001) for scorer 1, and 16.6 (16.7) vs 20 (18.8) (p < 0.001) for scorer 2. The sensitivity of Somté for a PSG-AHI > 5 was 91% for scorer 1 and 90% for scorer 2, while specificity was 77% and 90%, respectively. The areas under the receiver operating curve for different PSG-AHI cutoff points (≥ 5, ≥ 15, and ≥ 30) were 0.81, 0.90, and 0.86, respectively, for scorer 1, and 0.90, 0.88, and 0.83 for scorer 2.

CONCLUSIONS

These data suggest that Somté is an effective device to identify sleep and respiratory variables in patients with suspected OSA.

Для цитирования: Ferré Alex, Sampol Gabriel, Jurado Maria José, Cambrodi Roser, Lloberes Patricia, Romero Odile. Neurophysiological two-channel polysomnographic device in the diagnosis of sleep apnea.. Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine. 2012. DOI: 10.5664/jcsm.1770.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.