DOI 10.1046/j.1460-9592.2001.01696.x

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2001

Cross-stimulation during lead impedance monitoring.

Beaufort-Krol G C, Bink-Boelkens MTh, Nagelkerke D, Van Gelder B M

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

At the age of 4 years, a total cavopulmonary connection was performed in a boy with a complex congenital heart defect. On addition, a DDDR pacemaker was implanted for sick sinus syndrome. Atrial and ventricular leads were epicardially placed at the left atrium and left ventricle. At the age of 10 years, a new epicardial ventricular lead was placed because of malfunction of the existing lead. At the same operation the pulse generator was replaced by a Medtronic Kappa DR 731. After replacement, the boy experienced episodes of phrenic nerve stimulation associated with feelings of discomfort. Holter recordings revealed ventricular stimulation from the atrial stimulus for 2 consecutive beats. This phenomenon repeated exactly every 3 hours and was caused by the automatic lead impedance measurement that used a 5-V, 1-ms stimulus output.

Для цитирования: Beaufort-Krol G C, Bink-Boelkens MTh, Nagelkerke D, Van Gelder B M. Cross-stimulation during lead impedance monitoring.. Pacing and clinical electrophysiology : PACE. 2001. DOI: 10.1046/j.1460-9592.2001.01696.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.