Pulmonary Capillary Hemorrhage Induced by Acoustic Radiation Force Impulse Shear Wave Elastography in Ventilated Rats.

Miller Douglas L, Dong Zhihong, Dou Chunyan, Patterson Brandon, Raghavendran Krishnan

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Diagnostic ultrasound (DUS) imaging can induce pulmonary capillary hemorrhage (PCH), possibly related to the ultrasonic radiation surface pressure arising from reflection at the lung blood-air interfaces. Acoustic radiation force impulse (ARFI) elastography is a relatively new DUS mode with high-energy "push pulses" used to move tissue and generate shear waves. The objective of this study was to characterize PCH induced by the ARFI elastographic mode for comparison with other previously tested DUS modes.

METHODS

Pulmonary capillary hemorrhage induction was examined for ARFI elastographic frames with 5.7-MHz push pulses (Acuson S3000; Siemens Medical Solutions, Mountain View, CA), which had a derated PRPA of 2.6 MPa. Groups of rats with tracheal tube placement had no ventilation (spontaneous breathing), intermittent positive pressure ventilation (IPPV), or IPPV plus 8 cm H2 O of positive end-expiratory pressure (PEEP). Exposure was to 1 or 20 manually triggered image frame acquisitions. The PCH area was measured on the lung surface.

RESULTS

All 20-frame exposure groups, and even the single-frame group, had significant PCH relative to shams. Single-frame exposures produced significantly less PCH (P = .002) than 20-frame exposures in rats with a tracheal tube only (spontaneous breathing). The PEEP inhibited the PCH and produced about half of the PCH area induced for IPPV without PEEP (P = .014).

CONCLUSIONS

The PCH results were comparable with those from a previous study using B-mode or color Doppler exposure for 5 minutes; however, these modes delivered many more pulses for continuous imaging frames, suggesting that the ARFI elastographic frames were individually much more effective.

Для цитирования: Miller Douglas L, Dong Zhihong, Dou Chunyan, Patterson Brandon, Raghavendran Krishnan. Pulmonary Capillary Hemorrhage Induced by Acoustic Radiation Force Impulse Shear Wave Elastography in Ventilated Rats.. Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine. 2019. DOI: 10.1002/jum.14950.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.