BACKGROUND

Unrecognized hemorrhage and unguided resuscitation is associated with increased perioperative morbidity and mortality. The authors investigated peripheral venous waveform analysis (PIVA) as a method for quantitating hemorrhage as well as iatrogenic fluid overload during resuscitation.

METHODS

The authors conducted a prospective study on Yorkshire Pigs (n = 8) undergoing hemorrhage, autologous blood return, and administration of balanced crystalloid solution beyond euvolemia. Intra-arterial blood pressure, electrocardiogram, and pulse oximetry were applied to each subject. Peripheral venous pressure was measured continuously through an upper extremity standard peripheral IV catheter and analyzed with LabChart. The primary outcome was comparison of change in the first fundamental frequency (f1) of PIVA with standard and invasive monitoring and shock index (SI).

RESULTS

Hemorrhage, return to euvolemia, and iatrogenic fluid overload resulted in significantly non-zero slopes of f1 amplitude. There were no significant differences in heart rate or mean arterial pressure, and a late change in SI. For the detection of hypovolemia the PIVA f1 amplitude change generated an receiver operator curves (ROC) curve with an area under the curve (AUC) of 0.93; heart rate AUC = 0.61; mean arterial pressure AUC = 0.48, and SI AUC = 0.72. For hypervolemia the f1 amplitude generated an ROC curve with an AUC of 0.85, heart rate AUC = 0.62, mean arterial pressure AUC = 0.63, and SI AUC = 0.65.

CONCLUSIONS

In this study, PIVA demonstrated a greater sensitivity for detecting acute hemorrhage, return to euvolemia, and iatrogenic fluid overload compared with standard monitoring and SI. PIVA may provide a low-cost, minimally invasive monitoring solution for monitoring and resuscitating patients with perioperative hemorrhage.

Для цитирования: Hocking Kyle M, Sileshi Ban, Baudenbacher Franz J, Boyer Richard B, Kohorst Kelly L, Brophy Colleen M, Eagle Susan S. Peripheral Venous Waveform Analysis for Detecting Hemorrhage and Iatrogenic Volume Overload in a Porcine Model.. Shock (Augusta, Ga.). 2016. DOI: 10.1097/shk.0000000000000615.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.