MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/md.0000000000042899

Медицинская библиотека · Medicine · 2025

Validation of blood pressure measurement using a smartwatch in patients with acute ischemic stroke.

Liu Yu-Hsuan, Chen Li-Ching, Guo Yi-Shian, Lin Huan-Jan

СтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

We aimed to validate the accuracy of blood pressure (BP) measurement using a smartwatch in patients with acute ischemic stroke. We compared 140 pairs of BP (n = 35) measurements acquired by a smartwatch (ASUS VivoWatch SP) with those measured by a sphygmomanometer (reference device). Differences between the smartwatch BP and reference BP measurements were compared. The validation procedure and criterion followed the consensus of the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2018 and extended the standard to the specificities of cuffless devices in acute ischemic stroke population. The mean and standard deviation of the differences measured by a smartwatch and reference device were 1.8 ± 5.7 mm Hg in systolic BP and 0.7 ± 3.6 mm Hg in diastolic BP according to criterion 1. The mean and standard deviation of the differences measured by a smartwatch and reference device were 1.8 ± 5.6 mm Hg in systolic BP and 0.7 ± 3.6 mm Hg in diastolic BP according to criterion 2. The results both met the standards of the 2 criterion. The validation result did not differ between the paralytic and non-paralytic arms. The smartwatch with photoplethysmography sensors can provide accurate and reliable measurement of BP in acute ischemic stroke patients.

Для цитирования: Liu Yu-Hsuan, Chen Li-Ching, Guo Yi-Shian, Lin Huan-Jan. Validation of blood pressure measurement using a smartwatch in patients with acute ischemic stroke.. Medicine. 2025. DOI: 10.1097/md.0000000000042899.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.