Impact of x-ray contamination in the dose monitor on beam abortion during uninterrupted continuous delivery in proton therapy.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yagi Masashi, Lew Kah Seng, Chua Clifford Ghee Ann, Koh Wei Yang Calvin, Lee Kang Hao, Lee James Cheow Lei, Park Sung Yong, Tan Hong Qi
Аннотация
Objective.Real-time gated proton therapy (RGPT) enables precise irradiation of moving targets and is an important motion-management technique in proton therapy. Uninterrupted continuous delivery (UCD) avoids frequent gating interruptions, which improves proton beam extraction stability and treatment efficiency in synchrotron-based systems. However, because fluoroscopic x-rays remain on during proton irradiation in UCD, scattered fluoroscopic x-rays may be detected by the dose monitor (DM) and mistakenly counted as proton monitor units (MUs), potentially causing beam aborts. This study aimed to quantify scattered fluoroscopic x-rays detected by the DM during RGPT and evaluate their impact on beam delivery during simultaneous fluoroscopy and proton irradiation.Methods.Scattered fluoroscopic x-ray signals detected by the DM were measured using an RGPT system. Phantom experiments evaluated scattered x-ray dose rates using plastic water® (PW) and anthropomorphic phantoms while varying phantom thickness, fluoroscopic parameters, and range-shifter conditions. Beam-abort simulations were conducted under UCD conditions for both phantom and clinical liver treatment plans. Abort probability was assessed as a function of proton MU per spot, scattered x-ray MU per frame, and fluoroscopy frames per second.Results.DM signals from scattered fluoroscopic x-rays increased with tube voltage, tube current, field size, and phantom thickness, and were higher without a range shifter (RS). The maximum dose rate was 0.0211 MU/frame at 125 kV, 80 mA, 200 × 200 mm2, and PW-30 cm without a RS, representing an 11.7-fold increase compared to the in-air conditions. The results obtained from the anthropomorphic phantom experiments further validated the findings observed in the PW phantom study. Beam-abort simulations showed abort proportions below 2% for monoenergetic fields and below 3% for clinical liver plans across all tested conditions.Significance.With the DM positioned closer to the x-ray tube, scattered fluoroscopic x-ray contamination increased with phantom thickness, differing from prior reports. Optimization of proton dose rate, imaging parameters, and system geometry can minimize beam interruptions, supporting the clinical feasibility of UCD.