DOI 10.1097/brs.0000000000000507

Медицинская библиотека · Spine · 2014

Bioabsorbable interbody magnesium-polymer cage: degradation kinetics, biomechanical stiffness, and histological findings from an ovine cervical spine fusion model.

Daentzer Dorothea, Willbold Elmar, Kalla Katharina, Bartsch Ivonne, Masalha Waseem, Hallbaum Maximiliane, Hurschler Christof, Kauth Theresa, Kaltbeitzel Daniel, Hopmann Christian, Welke Bastian

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

STUDY

DESIGN

An experimental study using a sheep cervical spine interbody fusion model.

OBJECTIVE

First, to compare anterior cervical discectomy and fusion of an experimental bioabsorbable cage consisting of a magnesium alloy and a polymer (poly-ε-caprolactone, PCL) with an autologous tricortical iliac crest bone graft. Second, to determine the degradation kinetics of the cage, assess the 2 fusion devices for biomechanical stability, and determine their histological characteristics.

SUMMARY OF BACKGROUND DATA: Bioabsorbable cages are not routinely used in spine surgery at present, due to some undesirable effects such as cracks and foreign body reactions. This study involved the manufacture of a bioabsorbable cage from a magnesium alloy and the polymer PCL, which was then used as a device for anterior cervical discectomy and fusion in a sheep cervical spine fusion model.

METHODS

Twenty-four sheep had anterior cervical discectomy and fusion of C3-C4 and C5-C6 with an experimental bioabsorbable cage consisting of the magnesium alloy AZ31, which was infiltrated and covered with PCL at 1 level and with an autologous tricortical iliac crest bone graft at a second level. The sheep were divided into 4 groups. After 3, 6, 12, or 24 weeks postimplantation, the animals were killed and the cervical spines were harvested. The intervertebral spaces with the cage were investigated using μ-computed tomographic images to calculate degradation kinetics. Stiffness of all monosegments was determined through biomechanical testing. Histological analysis was performed to evaluate fusion status and to detect any foreign body reactions. The results from both implants were compared.

RESULTS

The magnesium-PCL cage showed nonlinear degradation over time. Both implants demonstrated time-dependent increases in stability, with a significantly greater stiffness of the bone graft after 24 weeks in all directions of motion. Histologically, the cage showed no signs of fusion with progressive encapsulation over time.

CONCLUSION

In comparison with the bone graft, the bioabsorbable cage showed inferior stiffness and fusion properties. Thus, further component modifications are necessary.

LEVEL OF EVIDENCE: N/A.

Для цитирования: Daentzer Dorothea, Willbold Elmar, Kalla Katharina, Bartsch Ivonne, Masalha Waseem, Hallbaum Maximiliane, Hurschler Christof, Kauth Theresa, Kaltbeitzel Daniel, Hopmann Christian, Welke Bastian. Bioabsorbable interbody magnesium-polymer cage: degradation kinetics, biomechanical stiffness, and histological findings from an ovine cervical spine fusion model.. Spine. 2014. DOI: 10.1097/brs.0000000000000507.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.