Safety and efficacy of the latest generation biodegradable polymer-coated ultrathin sirolimus-eluting stent in the treatment of coronary artery disease in a European all-comer population with or without high bleeding risk: The Cruz HBR Registry.

Leistner David M, Rampat Rajiv, Haude Michael, Schmitz Thomas, Allali Abdelhakim, Möllmann Helge, Stähli Barbara E, Rudolph Tanja K, Lauten Alexander, Koning René, Bogaerts Kris, Sudhir Krishnankutty, Naber Christoph

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The latest generation ultrathin Supraflex Cruz (Sahajanand Medical Technologies Limited, Surat, India) sirolimus-eluting stent (SES) has shown early healing properties and represents an attractive percutaneous coronary intervention (PCI) device in a high bleeding risk (HBR) population. The aim of this Cruz HBR registry was to assess safety and efficacy of the Supraflex Cruz SES in a large cohort of all-comer patients, of whom about one third were patients at HBR.

METHODS

Patients undergoing PCI were enrolled in this prospective, multi-centre, open label registry and stratified into non-HBR and HBR groups. The primary endpoint was a device-oriented composite endpoint (DOCE), a composite of cardiovascular death, myocardial infarction not clearly attributable to a non-target vessel and clinically driven target lesion revascularization within 12 months after PCI. The predefined aims were to show non-inferiority of the non-HBR group to the Supraflex arm of the TALENT Trial, and of the HBR group to polymer-free biolimus-coated stent arm of LEADERS FREE Trial.

RESULTS

A total of 1203 patients were enrolled across 26 European centers, including a significant proportion (38.7%; N.=466) of HBR patients. A total of 1745 lesions were treated in 1203 patients and 2235 stents were implanted. The DOCE occurred within the total cohort in 5.8% of patients with a significant difference between HBR patients and non-HBR patients (8.1% vs. 4.4%; P<0.001). All-cause mortality at 12 months was significantly (P<0.0001) different among HBR (9.0%) and non-HBR patients (1.7%), respectively. At 12 months, the overall incidence of definite and probable stent thrombosis was 1.0%. Major bleeding occurred in 5.9% patients of the HBR group. These results met the non-inferiority criteria with respect to the TALENT trial for the non-HBR group (P<0.0001), and the LEADERS FREE trial for the HBR group (P<0.0001).

CONCLUSIONS

The Cruz HBR registry confirms that PCI with the Supraflex Cruz SES is associated with a favorable clinical outcome in an all-comer population, including complex patients with HBR.

Для цитирования: Leistner David M, Rampat Rajiv, Haude Michael, Schmitz Thomas, Allali Abdelhakim, Möllmann Helge, Stähli Barbara E, Rudolph Tanja K, Lauten Alexander, Koning René, Bogaerts Kris, Sudhir Krishnankutty, Naber Christoph. Safety and efficacy of the latest generation biodegradable polymer-coated ultrathin sirolimus-eluting stent in the treatment of coronary artery disease in a European all-comer population with or without high bleeding risk: The Cruz HBR Registry.. Minerva cardiology and angiology. 2024. DOI: 10.23736/s2724-5683.24.06462-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.