DOI 10.1097/opx.0000000000000804

Медицинская библиотека · Optometry and vision science : official publication of the American Academy of Optometry · 2016

Rigid Gas Permeable Contact Lens Fitting Using New Software in Keratoconic Eyes.

Ortiz-Toquero Sara, Rodriguez Guadalupe, de Juan Victoria, Martin Raul

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine the repeatability of the back optic zone radius (BOZR) of rigid gas permeable (GP) contact lens (CL) proposed by new software for fitting in healthy and keratoconus eyes and to compare with the diagnostic CL fitting method.

METHODS

Three consecutive corneal topographies (Oculus-Keratograph) were performed and analyzed with APEX new software CL fitting (Hecht-Contactlinsen, Germany) in 40 healthy and 40 keratoconus eyes fitted with GP using conventional diagnostic method. The coefficient of variation (CV) of the BOZR suggested by APEX was calculated. The BOZR of both fitting methods (software versus diagnostic) were compared maintaining the same lens diameter.

RESULTS

BOZR proposed by APEX showed good repeatability in healthy (CV = 0.32%) and keratoconus eyes (CV = 0.51%). APEX proposed flatter BOZR than the diagnostic method in healthy (7.91 ± 0.24 and 7.84 ± 0.26 mm, p < 0.01) and keratoconus eyes (7.34 ± 0.38 and 7.23 ± 0.37 mm, p < 0.01). A strong linear correlation in healthy (BOZR_Diagnostic_Method = (BOZR_APEX × 1.06) - 0.53; p < 0.01, R = 0.969) and keratoconus eyes (BOZR_Diagnostic_Method = (BOZR_APEX × 0.88) + 0.77; p < 0.01, R = 0.825) was found. A detailed analysis showed a similar trend in different keratoconus stages (Amsler-Krumeich classification); stage 1: 7.42 ± 0.30 and 7.40 ± 0.25 mm, BOZR_Diagnostic_Method = (BOZR_APEX × 0.81) + 1.38, R = 0.973; stage 2: 7.30 ± 0.44 and 7.23 ± 0.38 mm, BOZR_Diagnostic_Method = (BOZR_APEX × 0.84) + 1.07, R = 0.929; and stage 3: 7.33 ± 0.39 and 7.11 ± 0.40 mm, BOZR_Diagnostic_Method = (BOZR_APEX × 0.93) + 0.28, R = 0.831. Applying these regression formulas, the BOZR difference could be reduced in healthy (-0.01 ± 0.05 mm) and keratoconus eyes (-0.01 ± 0.14 mm) for each keratoconus stage (0.01 ± 0.04, 0.03 ± 0.10, and 0.02 ± 0.16 mm in stages 1, 2, and 3, respectively).

CONCLUSIONS

APEX software provides repeatable BOZR in healthy and keratoconus eyes, but it tends to propose flatter BOZR than the diagnostic method. APEX BOZR should be improved with new equations and helping with the GP fitting procedure.

Для цитирования: Ortiz-Toquero Sara, Rodriguez Guadalupe, de Juan Victoria, Martin Raul. Rigid Gas Permeable Contact Lens Fitting Using New Software in Keratoconic Eyes.. Optometry and vision science : official publication of the American Academy of Optometry. 2016. DOI: 10.1097/opx.0000000000000804.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.