Tricuspid insufficiency after laser lead extraction.

Rodriguez Yasser, Mesa Julian, Arguelles Eric, Carrillo Roger G

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The use of laser lead extraction (LLE) to remove pacemaker and implantable cardiac defibrillator leads has become more prevalent in the past decade. Though the procedure is associated with a low rate of complications, LLE still poses some risks to patients. Some reports have suggested an increase in tricuspid insufficiency (TI) associated with LLE. We present a series of patients who underwent both LLE and complete evaluation for TI with echocardiographic techniques.

METHODS

From August 2008 to January 2010, 173 prospective, consecutive patients underwent LLE in a single center. All patients had transesophageal echocardiograms (TEE) during the extraction. Fifty-three patients had tricuspid valve function evaluated a day before the procedure with a transthoracic echocardiogram (TTE), during the procedure with a TEE and 2 days postoperatively with a TTE.

RESULTS

All 173 patients experienced no change in tricuspid valve function during the procedure with TEE. Of the 53 patients who underwent a complete TI evaluation, 38 were males (72%) and 15 females (38%), with a mean age of 69.45 ± 14.08. Mean ejection fraction was 35.82 ± 14.72. Three (6%) patients experienced TI after the procedure (two mild and one severe, all with tricuspid valve endocarditis); 16 (30%) patients were found to have TI before LLE that returned to normal valve function during or after the procedure. Thirty-four (64%) patients did not experience any significant change of the tricuspid valve performance after LLE.

CONCLUSION

LLE was not associated with increased TI.

Для цитирования: Rodriguez Yasser, Mesa Julian, Arguelles Eric, Carrillo Roger G. Tricuspid insufficiency after laser lead extraction.. Pacing and clinical electrophysiology : PACE. 2013. DOI: 10.1111/pace.12160.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.