DOI 10.1161/hypertensionaha.124.24640

Медицинская библиотека · Hypertension (Dallas, Tex. : 1979) · 2025

Sex Differences in BP From an Automated Oscillometric Compared With Manual Device.

Bui Tan V, Sharman James E, Chapman Niamh, Schultz Martin G, Ringrose Jennifer S, Gall Seana L, Bothe Tomas L, Brady Tammy M, Picone Dean S

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Automated blood pressure (BP) devices may be less accurate in females than males, but this requires further investigation. This study aimed to determine sex differences in automated BP, measured with a single brand and model of device, compared with manual BP, with a focus on cuff sizes and associations with measures of adiposity.

METHODS

Automated (Omron HEM-907XL) and manual BP were taken sequentially in a random order among a subsample of participants attending the US National Health and Nutrition Examination Survey, 2017 to 2018. Anthropometry and dual-energy x-ray absorptiometry were used to record body size and composition. Analyses, including multivariable regression to determine sex differences in BP, by cuff size, followed complex survey statistical principles.

RESULTS

A total of 3735 participants (49.0% female [95% CI, 46.4-51.6], 45 years [43-46]) were included. In females, automated systolic BP (SBP) incrementally underestimated manual SBP across larger cuffs up to extra-large (-6.4 mm Hg [-8.0 to -4.9]). In males, automated SBP underestimated manual SBP only with extra-large cuffs (-2.4 mm Hg [95% CI-3.9 to -0.9]). Underestimation by automated SBP with extra-large cuffs was independently associated with all measures of body size indicative of increased adiposity in both females and males. Hypertension classification from automated and manual SBP had moderate agreement for adult/large cuffs (weighted kappa range 0.66-0.79) but weak agreement for extra-large cuffs (0.55-0.58) for females and males.

CONCLUSIONS

The automated device used in this study underestimated manual SBP at larger cuff sizes, which was associated with indices of adiposity. Poorer accuracy of automated BP in larger cuff sizes could contribute to inequitable BP-related health care for females and males and requires further investigation.

Для цитирования: Bui Tan V, Sharman James E, Chapman Niamh, Schultz Martin G, Ringrose Jennifer S, Gall Seana L, Bothe Tomas L, Brady Tammy M, Picone Dean S. Sex Differences in BP From an Automated Oscillometric Compared With Manual Device.. Hypertension (Dallas, Tex. : 1979). 2025. DOI: 10.1161/hypertensionaha.124.24640.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.