MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/08164622.2023.2277287

Медицинская библиотека · Clinical & experimental optometry · 2023

Precision and agreement of axial length in paediatric population measured with MYAH and AL-Scan biometers.

Aldaba Mikel, Ochando Pablo, Vila-Vidal Nuria, Vinuela-Navarro Valldeflors, Guisasola Laura, Perez-Corral Joan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

CLINICAL RELEVANCE: Measuring axial length is key in the field of myopia development and control. Hence, the precision and agreement of commercially available biometers is of vital interest to understand their variability and interchangeability in the paediatric population.

BACKGROUND

Different biometers are available to measure axial length and monitor myopia progression in clinical practice. The purpose of this study was to assess the precision (repeatability and reproducibility) and agreement of the MYAH and AL-Scan biometers in a paediatric population.

METHODS

Three consecutive measurements were performed using MYAH and AL-Scan biometers in each subject by the same operator to test for repeatability. To test for reproducibility, two measurements were performed for each subject by two different observers with a 5-min interval between measurements. To test the agreement, each subject was measured once with each instrument.

RESULTS

A total of 187 subjects, with a mean age of 8.5 ± 0.3 years and mean spherical equivalent refractive error of +0.22 ± 0.77 D participated in the study. For the repeatability study, the within-subject standard deviation was 0.01 mm, and the repeatability limit was 0.04 mm for both instruments, with no statistically significant differences among repeated measures (p = 0.162 for MYAH and p = 0.774 for AL-Scan). For the reproducibility study, the within-subject standard deviation was 0.01 mm and the repeatability limit was 0.04 mm. There were statistically significant differences for the repeated measures for the AL-Scan (p = 0.002) but not for the MYAH (p = 0.643). Regarding the agreement between both instruments, the 95% limit of agreement ranged from -0.04 to 0.05 mm, and the differences were statistically significant (p = 0.021).

CONCLUSIONS

The repeatability, reproducibility, and agreement of the MYAH and AL-Scan biometers seem optimal for following children with myopia.

Для цитирования: Aldaba Mikel, Ochando Pablo, Vila-Vidal Nuria, Vinuela-Navarro Valldeflors, Guisasola Laura, Perez-Corral Joan. Precision and agreement of axial length in paediatric population measured with MYAH and AL-Scan biometers.. Clinical & experimental optometry. 2023. DOI: 10.1080/08164622.2023.2277287.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.