MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.mlg.0000165375.08563.18

Медицинская библиотека · The Laryngoscope · 2005

Steady-state response audiometry in a group of patients with steeply sloping sensorineural hearing loss.

Ballay Charles, Tonini Ross, Waninger Terra, Yoon Caroline, Manolidis Spiros

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The purpose of this study was to examine the predictive value of auditory steady-state response (ASSR) evoked potential thresholds and predicted behavioral thresholds in a group of children with steeply sloping sensorineural hearing loss (HL).

STUDY

DESIGN

Case series.

METHODS

Twenty-nine children with sloping sensorineural HL underwent behavioral audiometric evaluation, impedance testing, distortion product otoacoustic emissions, and steady-state response testing. A t test was performed to compare the means of ASSR predicted behavioral thresholds and behavioral responses. Pearson correlation coefficients were calculated at each tested frequency, 500 Hz, 1,000 Hz, 2,000 Hz, and 4,000 Hz, using the same data.

RESULTS

Bracketed thresholds were obtained at 500 Hz, 1,000 Hz, 2,000 Hz, and 4,000 Hz. Nineteen ears were used in this evaluation. A comparison of threshold difference as a function of bracketing revealed that the means were statistically different (P < .05). The mean threshold differences were calculated, and Pearson r values were determined between the behavioral thresholds and the predicted thresholds using the Rance 95 algorithm. The results revealed no difference of means at 500 Hz between predicted and measured behavioral thresholds. Linear regression analysis revealed strong correlation at 500 Hz, 1,000 Hz, and 2,000 Hz.

CONCLUSIONS

The GSI Audera appears to predict the configuration of HL in children with steeply sloping sensorineural HLs and over-predicts the severity of the loss by 15 to 20 dB above 500 Hz at each test frequency (1,000, 2,000, and 4,000 Hz). Correlation coefficients display a strong correlation at 500 Hz, 1,000 Hz, and 2,000 Hz.

Для цитирования: Ballay Charles, Tonini Ross, Waninger Terra, Yoon Caroline, Manolidis Spiros. Steady-state response audiometry in a group of patients with steeply sloping sensorineural hearing loss.. The Laryngoscope. 2005. DOI: 10.1097/01.mlg.0000165375.08563.18.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.