DOI 10.3881/j.issn.1000-503x.2016.02.003
[Effect of Paroxysmal Atrial Fibrillation on Parameters Monitored by Transpulmonary Thermodilution Method].
Аннотация
OBJECTIVE
To investigate the effect of atrial fibrillation on the accuracy of parameters monitored by transpulmonary thermodilution method.
METHODS
Totally 12 patients from emergency intensive care unit with paroxysmal atrial fibrillation were enrolled. The hemodynamic parameters such as heart rate, mean arterial pressure, cardiac index, systemic vascular resistance index, intrathoracic blood volume index, and extravascular lung water index were monitored by transpulmonary thermodilution method before paroxysmal atrial fibrillation and during atrial fibrillation, the number of B-lines was detected by lung ultrasonography before and during paroxysmal atrial fibrillation. The changes of all the parameters were analyzed.
RESULTS
When the paroxysmal atrial fibrillation happened, the heart rate increased significantly [(123.3±20.0) beat/min vs. (98.9±12.3) beat/min, P=0.006]; the mean arterial pressure [(86.9±10.2) mmHg vs. (93.0±12.5) mmHg, P=0.058], cardiac index [(2.82±0.62) L/(min·m(2)) vs. (3.31±1.02) L/(min·m(2)), P=0.058] and systemic vascular resistance index [(2254±947) dyn·s·cm(-5)·m(2) vs. (2302±828) dyn·s·cm(-5)·m(2), P=0.351] had no obvious change; however, the intrathoracic blood volume index significantly increased [(1333±90) ml/m(2) vs. (937±111) ml/m(2), P<0.001]; extravascular lung water index also increased significantly [(16.1±1.1) ml/kg vs. (6.5±1.9) ml/kg, P<0.001]. No significant difference was found in the number of B-lines detected by lung ultrasonography before and during atrial fibrillation (10.0±4.2 vs. 9.4±4.4, P=0.180).
CONCLUSION
Both intrathoracic blood volume and extravascular lung water monitored by transpulmonary thermodilution method were overvalued during paroxysmal atrial fibrillation, which may mislead the clinical judgment and decision-making.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.