Bench evaluation of MyVal OCTACOR THVs at expansion limits for paediatric and valve-in-valve applications.
СтатьяПеревод в очереди
Аннотация
Balloon-expandable MyVal OCTACOR valves were evaluated in 3D-printed conduits under simulated flow. The 20 mm valve functioned from 16-24 mm, rupturing at 27 mm. The 29 mm expanded to 33 mm with mild insufficiency. Findings suggest OCTACOR valves may support repeated valve-in-valve procedures, with the 20 mm showing greater versatility for small-infant mitral replacements and as a potential alternative to Melody valves.
Для цитирования: Haddad Raymond N, Ciobotaru Vlad, Hascoët Sébastien, Fraisse Alain. Bench evaluation of MyVal OCTACOR THVs at expansion limits for paediatric and valve-in-valve applications.. Cardiology in the young. 2025. DOI: 10.1017/s104795112510992x.