MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1515/jpem.2011.420

Медицинская библиотека · Journal of pediatric endocrinology & metabolism : JPEM · 2011

Validation of automatic bone age rating in children with precocious and early puberty.

Martin David D, Meister Kathrin, Schweizer Roland, Ranke Michael B, Thodberg Hans Henrik, Binder Gerhard

СтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIMS: Manual bone age (BA) rating in precocious puberty (PP) is associated with considerable rater variability. The aim was to evaluate a new method for automated Greulich and Pyle (GP) BA determination in children with PP.

METHODS

Seven hundred forty-one archived X-rays from 13 boys and 103 girls with PP or early puberty of various etiologies (age range at time of X-ray, 0.3-14.8 years; mean BA advancement, 2.3 years) were rated. Automatic rating (BoneXpert BA, or BXBA) was compared with the original manual GP rating (manual BA, or ManBA). X-rays where BXBA deviated from ManBA by more than 1.5 years were rerated by three raters, and the average was formed (ReferenceBA).

RESULTS

All 741 X-rays, except nine (three images had poor quality and six were from children with a chronological age younger than 1.5 years), were analyzed automatically. The mean difference of BXBA-ManBA was -0.19 years; the SD of the differences was 0.76 years (95% confidence interval 0.72-0.80). ReferenceBA was determined for 41 images. A discrepancy from ReferenceBA greater 1.5 years was found in four images against BXBA and in 10 images against ManBA.

CONCLUSION

Automated BA is efficient and reliable in children with PP.

Для цитирования: Martin David D, Meister Kathrin, Schweizer Roland, Ranke Michael B, Thodberg Hans Henrik, Binder Gerhard. Validation of automatic bone age rating in children with precocious and early puberty.. Journal of pediatric endocrinology & metabolism : JPEM. 2011. DOI: 10.1515/jpem.2011.420.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.