Медицинская библиотека · The Malaysian journal of pathology · 2023

Evaluation of quantitative point-of-care test for measurement of glucose-6-phosphate dehydrogenase enzyme activity in Malaysia.

Zailani M A H, Raja Sabudin R Z A, Abdul Jalil D, Ithnin A, Ayub N A, Alauddin H, Jalil N, Md Fauzi A, Cheah F C, Lim L S, Safian N, Yusuf M M, Othman A

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

The treatment of Plasmodium vivax malaria with 8-aminoquinolines is contraindicated in glucose-6-phosphate dehydrogenase (G6PD) deficient individuals due to the risk of acute haemolytic anaemia. Effective G6PD screening is paramount to avoid adverse drug reactions. This study aimed to evaluate the performance of novel quantitative point-of-care (POC) tests as a new screening method for G6PD deficiency in Malaysia.

MATERIALS AND

METHODS

A total of 153 neonatal cord blood, 99 peripheral blood of older children aged between 1 month to 12-years old, and 62 peripheral adult blood were screened for G6PD deficiency using two quantitative POC tests, CareStartTM biosensor (Carestart) and CareStartTM Biosensor 1 (S1). The results were compared with OSMMR2000D kit as a reference assay. Two statistical analyses were performed in this study to evaluate the POC test performances, the Spearman's correlation test and the Cohen's kappa method.

RESULTS

Both Carestart and S1 tests showed significant positive correlations to OSMMRS000D with r2 = 0.7916 and r2 = 0.7467. Their measurement of agreement showed a kappa (κ) value of 0.805 (p<0.001, 95% CI), and 0.795 (p<0.001, 95% CI), respectively. Analysis of the area under the Receiver Operating Curve (ROC) at 60% cut-off illustrated that the Carestart had 90.2% sensitivity, 98.9% specificity, 98.3% positive predictive value (PPV), and 93.8% negative predictive value (NPV). The corresponding values for the S1 were 95.2%, 100%, 100%, and 96.8%, respectively.

CONCLUSION

This study showed that the Carestart and S1 biosensors have high-performance reliability for screening of G6PD deficiency, which can guide safe prescriptions of anti-malaria medications and hence, eradication of Plasmodium vivax malaria.

Для цитирования: Zailani M A H, Raja Sabudin R Z A, Abdul Jalil D, Ithnin A, Ayub N A, Alauddin H, Jalil N, Md Fauzi A, Cheah F C, Lim L S, Safian N, Yusuf M M, Othman A. Evaluation of quantitative point-of-care test for measurement of glucose-6-phosphate dehydrogenase enzyme activity in Malaysia.. The Malaysian journal of pathology. 2023.