DOI 10.1097/ico.0000000000000476
Distribution and Repeatability of Corneal Astigmatism Measurements (Magnitude and Axis) Evaluated With Color Light Emitting Diode Reflection Topography.
Аннотация
PURPOSE: To evaluate and investigate the distribution and repeatability of anterior corneal surface astigmatism measurements (axis and magnitude) using a novel corneal topographer.
METHODS
Anterior corneal surface astigmatism was investigated in a total of 195 eyes using a novel multicolored spot reflection topographer (Cassini; i-Optics). Two patient groups were studied, a younger-age group A and an older-age group B. Three consecutive acquisitions were obtained from each eye. The repeatability of measurement was assessed using Bland-Altman plot analysis and is reported as the coefficient of repeatability.
RESULTS
Group A (average age 34.3 years) had on average with-the-rule astigmatism, whereas the older-age group B (average age 72.3 years) had on average against-the-rule astigmatism. Average astigmatism magnitude measurement repeatability in group A was 0.4 diopters (D) and in group B 0.4 D. Average astigmatism axis measurement repeatability in group A was 5.4 degrees and in group B 5.5 degrees. The axis measurement repeatability improved with increasing magnitude of astigmatism: in the subgroups with astigmatism between 3.0 and 6.0 D, the axis repeatability was 1.4 degrees (group A) and 1.2 degrees (group B), whereas in the subgroups with astigmatism larger than 6.0 D, the repeatability was 1.1 and 0.6 degrees, respectively.
CONCLUSIONS
This novel corneal topography device seems to offer high precision in reporting corneal astigmatism. This study reaffirms the established trend of a corneal astigmatism shift from an average "with-the-rule" to "against-the-rule" with aging.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.