DOI 10.1016/j.jcrs.2014.06.029

Медицинская библиотека · Journal of cataract and refractive surgery · 2015

Comparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selection.

Srivannaboon Sabong, Chirapapaisan Chareenun, Chonpimai Pratuangsri, Koodkaew Sunisa

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare corneal astigmatism measurements obtained using a new optical biometer (AL-Scan) and a standard biometer based on partial coherence interferometry (IOLMaster 500) for toric intraocular lens (IOL) selection.

SETTING

Siriraj Hospital, Mahidol University, Bangkok, Thailand.

DESIGN

Prospective comparative study.

METHODS

Keratometry (K) readings were taken with the 2 biometers in eyes of cataract patients. The corneal astigmatism readings from the 2 devices were compared. The cylinder measurements were transformed into J0 and J45 vector components for analysis, and the devices' values were compared using the paired t test. Correlation coefficients were calculated. The Bland-Altman method was used to evaluate the agreement between the sets of measurements.

RESULTS

One hundred thirty-seven eyes (81 patients) were enrolled. The magnitude of astigmatism between the new biometer 2.4 mm zone and the standard biometer was similar (P = .19). There was a statistically significant difference between values from the new biometer 3.3 mm zone and the standard biometer (P < .05). There was good correlation in the overall magnitude of astigmatism between devices. There was a trend of low correlation toward the flat cornea (K <43.0 diopters [D]), low corneal cylinder (<1.0 D), and with-the-rule astigmatism. The IOL toricity calculated from both devices was similar.

CONCLUSIONS

Corneal astigmatism measurements with the 2 biometers were similar in magnitude, especially in 2.4 mm zone of the new biometer. There was no difference in the toric IOL selections between the devices.

FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.

Для цитирования: Srivannaboon Sabong, Chirapapaisan Chareenun, Chonpimai Pratuangsri, Koodkaew Sunisa. Comparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selection.. Journal of cataract and refractive surgery. 2015. DOI: 10.1016/j.jcrs.2014.06.029.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.