MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1540-8159.1992.tb03132.x

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 1992

Ventricular lead transection and atrial lead damage in a young softball player shortly after the insertion of a permanent pacemaker.

Schuger C D, Mittleman R, Habbal B, Wagshal A, Huang S K

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

We report a case in which permanent pacemaker implantation using a conventional subclavian approach on the throwing side of an avid softball player resulted in complete transection of the ventricular lead and severe damage to the atrial lead. The site of the lead fracture suggested that both leads were crushed between the clavicle and the first rib as a result of the frequent and repetitive arm movement. This case illustrates the importance of the selection of the correct approach for permanent pacing lead insertion.

Для цитирования: Schuger C D, Mittleman R, Habbal B, Wagshal A, Huang S K. Ventricular lead transection and atrial lead damage in a young softball player shortly after the insertion of a permanent pacemaker.. Pacing and clinical electrophysiology : PACE. 1992. DOI: 10.1111/j.1540-8159.1992.tb03132.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.