DOI 10.1093/bjr/tqae089

Медицинская библиотека · The British journal of radiology · 2024

Estimating brain and eye lens dose for the cardiologist in interventional cardiology-are the dose levels of concern?

Hulthén Markus, Tsapaki Virginia, Karambatsakidou Angeliki

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To establish conversion coefficients (CCs), between mean absorbed dose to the brain and eye lens of the cardiologist and the air kerma-area product, PKA, for a set of projections in cardiac interventional procedures. Furthermore, by taking clinical data into account, a method to estimate the doses per procedure, or annual dose, is presented.

METHODS

Thermoluminescence dosimeters were used together with anthropomorphic phantoms, simulating a cardiologist performing an interventional cardiac procedure, to estimate the CCs for the brain and eye lens dose for nine standard projections, and change in patient size and x-ray spectrum. Additionally, a single CC has been estimated, accounting for each projections fraction of use in the clinic and associated PKA using clinical data from the dose monitoring system in our hospital.

RESULTS

The maximum CCs for the eye lens and segment of the brain, is 5.47 μGy/Gycm2 (left eye lens) and 1.71 μGy/Gycm2 (left brain segment). The corresponding weighted CCs: are 3.39 μGy/Gycm2 and 0.89 μGy/Gycm2, respectively.

CONCLUSIONS

Conversion coefficients have been established under actual scatter conditions, showing higher doses on the left side of the operator. Using modern interventional x-ray equipment, interventional cardiac procedures will not cause high radiation dose levels to the operator when a ceiling mounted shield is used, otherwise there is a risk that the threshold dose values for cataract will be reached.

ADVANCE IN KNOWLEDGE: In addition to the CCs for the different projections, methods for deriving a single CC per cardiac interventional procedure and dose per year were introduced.

Для цитирования: Hulthén Markus, Tsapaki Virginia, Karambatsakidou Angeliki. Estimating brain and eye lens dose for the cardiologist in interventional cardiology-are the dose levels of concern?. The British journal of radiology. 2024. DOI: 10.1093/bjr/tqae089.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.