Comparison of transducers with different frequencies in breast contrast-enhanced ultrasound (CEUS) using SonoVue as contrast agent.

Wang Yong-Mei, Fan Wei, Zhang Kai, Zhang Li, Tan Zhen, Ma Rong

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To explore the effectiveness of different transducers in breast contrast-enhanced ultrasound (CEUS) using SonoVue(®) (Bracco, Plan-Les-Ouates, Switzerland) as the contrast agent.

METHODS

Breast CEUS was performed in 51 patients with 51 breast lesions using a low-frequency transducer (probe C5-1) and a high-frequency transducer (probe L12-5) separately. All image processes were reviewed for the presence of local blood perfusion defects and surrounding vessels. McNemar's test was conducted to compare the detection effectiveness between these two transducers.

RESULTS

Pathological results revealed 38 malignant and 13 benign lesions. The two transducers showed no difference in detecting benign lesions. Among malignant lesions, CEUS conducted by probe C5-1 (frequency range from 1 to 5 MHz) presented 23 (60.5%) lesions with local blood perfusion defects and 26 (68.4%) lesions with surrounding vessels. Meanwhile, probe L12-5 (frequency range from 5 to 12 MHz) showed only 12 (31.6%) lesions with local blood perfusion defects and 12 (31.6%) lesions with surrounding vessel. Probe C5-1 was more sensitive than probe L12-5 in detecting malignant CEUS characteristics (p-value < 0.05).

CONCLUSION

The low-frequency transducer was more sensitive than the high-frequency transducer in breast CEUS using SonoVue as the contrast agent. A new contrast agent with a higher resonance frequency, specially designed for high-frequency transducers, may be helpful in improving the clinical value of breast CEUS.

ADVANCES IN KNOWLEDGE: The first study comparing different frequency transducers in breast CEUS of the same patient lesions. We brought out the requirement for CEUS contrast agents which are more suitable for high-frequency examinations.

Для цитирования: Wang Yong-Mei, Fan Wei, Zhang Kai, Zhang Li, Tan Zhen, Ma Rong. Comparison of transducers with different frequencies in breast contrast-enhanced ultrasound (CEUS) using SonoVue as contrast agent.. The British journal of radiology. 2016. DOI: 10.1259/bjr.20151050.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.