m+Medelevator.RU
← Поиск публикаций
Pacing and clinical electrophysiology : PACE · 2013

Radiofrequency ablation at low irrigation flow rates using a novel 12-hole gold open-irrigation catheter.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Akca Ferdi, Zima Endre, Végh Eszter M, Széplaki Gábor, Skopál Judit, Hubay Martha, Lendvai Zsuzsanna, Merkely Bela, Szili-Torok Tamas

Аннотация

BACKGROUND: High irrigation rates during radiofrequency (RF) ablation may cause fluid overload and limit lesion size. This in vivo animal study assessed the safety and efficacy of RF ablation at low irrigation rates using a novel 12-hole gold catheter.

METHODS: A total of 103 lesions, created on the thigh of five mongrel dogs, were analyzed. Lesions were created using a 12-hole irrigated gold-tip (Au) and a six-hole irrigated platinum-iridium (PtIr) catheter (both 7F/3.5-mm electrode; BIOTRONIK SE & CO, KG, Berlin, Germany) in parallel and perpendicular orientation. RF current was delivered for 60 seconds at 30 W using 8 mL/min and 15 mL/min irrigation. Electrode temperature, steam pops, lesion dimensions, and coagulum formation were recorded.

RESULTS: Electrode temperatures were lower for Au compared to PtIr in parallel (8 mL/min: 38.1 ± 1.7°C vs 48.0 ± 4.8°C, P < 0.0001; 15 mL/min: 36.0 ± 1.5°C vs 46.9 ± 5.4°C, P < 0.0001) and perpendicular position (15 mL/min: 35.5 ± 1.2°C vs 38.4 ± 2.5°C, P = 0.003). The number of steam pops between Au and PtIr was comparable for parallel (8 mL/min: 14% vs 27%, P = 0.65; 15 mL/min: 14% vs 43%, P = 0.21) and perpendicular orientation (8 mL/min: 25% vs 17%, P = 1.00; 15 mL/min: 18% vs 0%, P = 0.48). Au created larger volumes than PtIr at 8 mL/min irrigation (861 ± 251 mm(3) vs 504 ± 212 mm(3) , P = 0.004); however, for 15 mL/min, volumes were comparable (624 ± 269 mm(3) vs 768 ± 466 mm(3) , P = 0.46). No coagulum formation was observed for any of the catheters on the surface and catheter tip.

CONCLUSION: RF ablation at low flow rate using a novel 12-hole irrigation Au catheter is safe and results in larger lesions than with a PtIr electrode.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты