Retrospective Comparison of the Myopia Master and the Lenstar LS900 Axial Length Measurements in Children with Myopia.

Sivakumar Thibika, Palmowski-Wolfe Anja

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study is a retrospective analysis to compare ocular biometry measurements of axial length in children with myopia using Myopia Master (OCULUS, Wetzlar, Germany) and Lenstar LS900 (HAAG-STREIT AG, Köniz, Switzerland).

PATIENTS AND

METHODS

Axial length measurements obtained with both instruments within a 3-week period were collected retrospectively. Measurements were visualized with a Bland-Altman plot. For statistical evaluation, a paired t-test was applied, and the Pearson correlation coefficient (r) was established.

RESULTS

Sixty-one eyes from 31 myopic patients (59% male, 41% female) with a mean age of 11.34 ± 3.25 years (range: 6 - 18 years) were identified. Mean axial length was 24.7 mm (SD 1.29) with the Myopia Master and 24.69 mm (SD 1.30) with the Lenstar LS900 (Pearson correlation: r = 0.9991). The average difference of the axial length measurement between the two biometers was 0.00 064 mm ± 0.056 SD (p = 0.9293).

CONCLUSION

The axial length measured by Myopia Master and Lenstar LS900 did not differ significantly. Thus, previous values obtained with the Lenstar LS900 can be applied to assess myopia progression.

Для цитирования: Sivakumar Thibika, Palmowski-Wolfe Anja. Retrospective Comparison of the Myopia Master and the Lenstar LS900 Axial Length Measurements in Children with Myopia.. Klinische Monatsblatter fur Augenheilkunde. 2023. DOI: 10.1055/a-2013-2557.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.