MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/00015385.2025.2459453

Медицинская библиотека · Acta cardiologica · 2025

Fifteen years of experience with the melody™ TPV for percutaneous pulmonary valve replacement.

Bjorn Cools, De Meester Pieter, Budts Werner, Heying Ruth, Vande Bruaene Alexander, Boshoff Derize, Depypere Anouk, Brown Stephen, Gewillig Marc

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

The Melody™ TPV has been used as an alternative to surgical pulmonary valve replacement; limited medium-term follow-up data are available.

AIMS: To report the follow-up data of all Melody™ TPVs implanted locally over a 15-year period (2006-2021).

METHODS

Single-centre non-randomised prospective observational study of all implanted Melody™ valves in the pulmonary position.

RESULTS

234 Melody™ valves were implanted at a mean age of 20.8 ± 24.6y. Indications for valve implantation included: pulmonary stenosis (47.2%,) regurgitation (30.9%), and mixed pathology (21.9%). The implant zone substrate consisted of homograft in 52.6%, patched right ventricular outflow tract in 33.8%, and bioprostheses in 13.6% of the cases. Valve survival at 10 years was 89% and 72% at 15 years follow-up. Pulmonary stenosis and pulmonary and tricuspid valve regurgitation demonstrated no significant evolution over the 15-year follow-up. Over the study period, there were 7 deaths at a mean age of 54.2 ± 21.1y; none was valve related. Valve failure was observed in 22 cases (9.4%), mainly due to endocarditis 13/22 (59.0%). The overall incidence of endocarditis was 1.5% per patient-year and occurred in 10.2% (n = 24) of patients 2.7 ± 1.6y after TPV, mostly in younger men (median 18.3, range 8.1 - 49.5 y). Balloon dilatation to accommodate for somatic growth was successful in all 17 (7.3%) attempted cases.

CONCLUSION

The Melody™ valve had a low risk for valve failure with overall well-preserved valve function over up to 15 years of follow-up. Endocarditis remains a concern. The Melody™ valve is competitive with other surgical and percutaneous conduits.

Для цитирования: Bjorn Cools, De Meester Pieter, Budts Werner, Heying Ruth, Vande Bruaene Alexander, Boshoff Derize, Depypere Anouk, Brown Stephen, Gewillig Marc. Fifteen years of experience with the melody™ TPV for percutaneous pulmonary valve replacement.. Acta cardiologica. 2025. DOI: 10.1080/00015385.2025.2459453.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.