Feasability of human endobronchial imaging with a linear-array ultrasound catheter.

Tranquart François, Palanchon Peggy, Cladé Odile, Bleuzen Aurore, Dinet Dominique, Magro Pascal, Diot Patrice

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To explore the feasability of imaging lung masses with a novel endobronchial linear-array transducer.

METHOD: We used a linear-array transducer of 7 F (2.3 mm in diameter) operating at a center frequency of 10 MHz for endobronchial imaging. We used the probe in 20 subjects with suspected lung cancer to identify and measure the tumor; to asses the tumor's sonographic characteristics, bronchial wall invasion, and presence of lymph nodes; to guide a transbronchial needle aspiration (TBNA); and to evaluate potential side effects.

RESULTS

In all patients, the tumor was identified and the relationship with the bronchial wall correctly evaluated through the depiction of a 3-layer wall pattern. Bronchial wall invasion was suspected in 5 patients, small adjacent lymph nodes were detected in 3 patients, and a small pleural effusion was observed close to the lung lesion in 2 patients. The linear-array probe allowed the guidance of successful TBNA with no complications.

CONCLUSION

This feasibility study shows that the prototype probe can be used to depict pulmonary lesions and to guide biopsy nodes. Larger series are needed to validate its usefulness in clinical work-ups and patients management.

Для цитирования: Tranquart François, Palanchon Peggy, Cladé Odile, Bleuzen Aurore, Dinet Dominique, Magro Pascal, Diot Patrice. Feasability of human endobronchial imaging with a linear-array ultrasound catheter.. Journal of clinical ultrasound : JCU. 2008. DOI: 10.1002/jcu.20512.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.