Validity and reproducibility of hand-held dynamometry in children aged 4-11 years.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
van den Beld Willeke A, van der Sanden Gitty A C, Sengers Rob C A, Verbeek André L M, Gabreëls Fons J M
Аннотация
OBJECTIVE: To evaluate validity and reproducibility of hand-held dynamometry in 11 different muscle groups in children.
DESIGN AND PATIENTS: Maximum isometric muscle strength was measured with a calibrated hand-held dynamometer in 61 patients aged 4-11 years who had been referred to our specialist centre in the past 3 years because of suspected myopathy. All the patients had had muscle biopsy.
METHODS: Validity was assessed by the power to discriminate between patients with and without myopathy, using logistic regression analysis and receiver operating characteristic analysis and sensitivity and specificity at a specifically chosen cut-off point. Reproducibility was evaluated by test-retest reliability in a stratified random sample of 40 patients who returned for re-measurements, using the intraclass correlation coefficient and the standard error of measurement.
RESULTS: In the patients, areas under the receiver operating characteristic curve ranged from 0.66 to 0.88. At a specifically chosen cut-off point, sensitivity varied from 73% to 87%, while specificity varied from 54% to 80%. Intraclass correlation coefficients ranged from 0.73 to 0.91. The standard error of measurement ranged from 3.3 N to 12.2 N.
CONCLUSION: Performance of hand-held dynamometry varied widely in the 11 muscle groups. Highest performance was observed in the elbow flexors. Test-retest reliability of the mean value of 2 efforts was generally higher than the maximum value.
DESIGN AND PATIENTS: Maximum isometric muscle strength was measured with a calibrated hand-held dynamometer in 61 patients aged 4-11 years who had been referred to our specialist centre in the past 3 years because of suspected myopathy. All the patients had had muscle biopsy.
METHODS: Validity was assessed by the power to discriminate between patients with and without myopathy, using logistic regression analysis and receiver operating characteristic analysis and sensitivity and specificity at a specifically chosen cut-off point. Reproducibility was evaluated by test-retest reliability in a stratified random sample of 40 patients who returned for re-measurements, using the intraclass correlation coefficient and the standard error of measurement.
RESULTS: In the patients, areas under the receiver operating characteristic curve ranged from 0.66 to 0.88. At a specifically chosen cut-off point, sensitivity varied from 73% to 87%, while specificity varied from 54% to 80%. Intraclass correlation coefficients ranged from 0.73 to 0.91. The standard error of measurement ranged from 3.3 N to 12.2 N.
CONCLUSION: Performance of hand-held dynamometry varied widely in the 11 muscle groups. Highest performance was observed in the elbow flexors. Test-retest reliability of the mean value of 2 efforts was generally higher than the maximum value.