Investigation of organs dosimetry precision using ATOM phantom and optically stimulated luminescence detectors in computed tomography.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Khallouqi Abdellah, Sekkat Hamza, El Rhazouani Omar, Halimi Abdellah
Аннотация
The primary objective of this study was to compare organ doses measured using optically stimulated luminescent dosimeters (OSLDs) with those estimated by the CT-EXPO software for common CT protocols. An anthropomorphic ATOM phantom was employed to measure organ doses across head, chest, and abdominal CT scans performed on a Hitachi Supria 16-slice CT scanner. These OSLD measurements were then compared to the estimates provided by the widely used CT-EXPO software. Organ doses were assessed using OSLDs placed in an adult anthropomorphic phantom, with calibration performed through a comprehensive process involving multiple tube potentials and sensitivity corrections. Results from three CT acquisitions per protocol were compared to estimates provided by CT-EXPO software. Findings reveal significant discrepancies between measured and estimated organ doses, with p-values consistently below 0.05 across all organs. For head CT, measured eye lens doses averaged 33.51 mGy, 6.0% lower than the estimated 35.65 mGy. In chest CT, the thyroid dose was 9.82 mGy, 13.5% higher than the estimated 8.65 mGy. For abdominal CT, the liver dose measured 12.11 mGy, 9.6% higher than the estimated 11.05 mGy. Measured doses for the rest of organs were generally lower than those predicted by CT-EXPO, showing some limitations in current estimation models and the importance of precise dosimetry. This study highlights the potential of OSLD measurements as a complementary method for organ dose assessment in CT imaging, emphasizing the need for more accurate organ dose measurement to optimize patient care.