MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1167/iovs.09-4403

Медицинская библиотека · Investigative ophthalmology & visual science · 2010

Optimizing hand-held spectral domain optical coherence tomography imaging for neonates, infants, and children.

Maldonado Ramiro S, Izatt Joseph A, Sarin Neeru, Wallace David K, Freedman Sharon, Cotten C Michael, Toth Cynthia A

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To describe age-related considerations and methods to improve hand-held spectral domain optical coherence tomography (HH-SD OCT) imaging of eyes of neonates, infants, and children.

METHODS

Based on calculated optical parameters for neonatal and infant eyes, individualized SD OCT scan parameters were developed for improved imaging in pediatric eyes. Forty-two subjects from 31 weeks postmenstrual age to 1.5 years were imaged with a portable HH-SD OCT system. Images were analyzed for quality, field of scan, magnification, and potential clinical utility.

RESULTS

The axial length of the premature infant eye increases rapidly in a linear pattern during the neonatal period and slows progressively with age. Refractive error shifts from mild myopia in neonates to mild hyperopia in infants. These factors affect magnification and field of view of optical diagnostic tools applied to the infant eye. When SD OCT parameters were corrected based on age-related optical parameters, SD OCT image quality improved in young infants. The field of scan and ease of operation also improved, and the optic nerve, fovea, and posterior pole were successfully imaged in 74% and 87% of individual eye imaging sessions in the intensive care nursery and clinic, respectively. No adverse events were reported.

CONCLUSIONS

SD OCT in young children and neonates should be customized for the unique optical parameters of the infant eye. This customization, not only improves image quality, but also allows control of the density of the optical sampling directed onto the retina.

Для цитирования: Maldonado Ramiro S, Izatt Joseph A, Sarin Neeru, Wallace David K, Freedman Sharon, Cotten C Michael, Toth Cynthia A. Optimizing hand-held spectral domain optical coherence tomography imaging for neonates, infants, and children.. Investigative ophthalmology & visual science. 2010. DOI: 10.1167/iovs.09-4403.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.