Pacing and clinical electrophysiology : PACE · 2014

Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement-a case-matched comparison in 198 patients.

Контролируемое клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Sigmund Elisabeth, Puererfellner Helmut, Derndorfer Michael, Kollias Georgios, Winter Siegmund, Aichinger Josef, Nesser Hans-Joachim, Martinek Martin

Аннотация

BACKGROUND: Sufficient electrode-tissue contact is crucial for adequate lesion formation in radiofrequency catheter ablation (RFCA).

OBJECTIVE: We assessed the impact of direct catheter force measurement on acute procedural parameters and outcome of RFCA for paroxysmal and persistent atrial fibrillation (AF).

METHODS: Ninety-nine consecutive patients (70% men) with paroxysmal (63.6%) or persistent AF underwent left atrial RFCA using a 3.5-mm open-irrigated-tip (OIT) catheter with contact force measurement capabilities (group 1). For comparison a case-matched cohort with standard OIT catheters was used (99 patients; group 2). Case matching included gender, type of AF, number or RFCA procedures, and type of procedure.

RESULTS: Procedural data showed a significant decline in radiofrequency ablation time from 52 ± 20 to 44 ± 16 minutes (P = 0.003) with a remarkable mean reduction in overall procedure time of 34 minutes (P = 0.0001; 225.8 ± 53.1 vs 191.9 ± 53.3 minutes). In parallel, the total fluoroscopy time could be significantly reduced from 28.5 ± 11.0 to 19.9 ± 9.3 minutes (P = 0.0001) as well as fluoroscopy dose from 74.1 ± 58.0 to 56.7 ± 38.9 Gy/cm(2) (P = 0.016). Periprocedural complications were similar in both groups.

CONCLUSIONS: The use of contact force sensing technology is able to significantly reduce ablation, procedure, and fluoroscopy times as well as dose in RFCA of AF in a mixed case-matched group of paroxysmal and persistent AF. Energy delivery is substantially reduced by avoiding radiofrequency ablation in positions with insufficient surface contact. Additionally 12-month outcome data showed increased efficacy. Such time saving and equally safe technology may have a relevant impact on laboratory management and increased cost effectiveness.