DOI 10.1111/pace.12549

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2014

Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement-a case-matched comparison in 198 patients.

Sigmund Elisabeth, Puererfellner Helmut, Derndorfer Michael, Kollias Georgios, Winter Siegmund, Aichinger Josef, Nesser Hans-Joachim, Martinek Martin

Контролируемое клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Sufficient electrode-tissue contact is crucial for adequate lesion formation in radiofrequency catheter ablation (RFCA).

OBJECTIVE

We assessed the impact of direct catheter force measurement on acute procedural parameters and outcome of RFCA for paroxysmal and persistent atrial fibrillation (AF).

METHODS

Ninety-nine consecutive patients (70% men) with paroxysmal (63.6%) or persistent AF underwent left atrial RFCA using a 3.5-mm open-irrigated-tip (OIT) catheter with contact force measurement capabilities (group 1). For comparison a case-matched cohort with standard OIT catheters was used (99 patients; group 2). Case matching included gender, type of AF, number or RFCA procedures, and type of procedure.

RESULTS

Procedural data showed a significant decline in radiofrequency ablation time from 52 ± 20 to 44 ± 16 minutes (P = 0.003) with a remarkable mean reduction in overall procedure time of 34 minutes (P = 0.0001; 225.8 ± 53.1 vs 191.9 ± 53.3 minutes). In parallel, the total fluoroscopy time could be significantly reduced from 28.5 ± 11.0 to 19.9 ± 9.3 minutes (P = 0.0001) as well as fluoroscopy dose from 74.1 ± 58.0 to 56.7 ± 38.9 Gy/cm(2) (P = 0.016). Periprocedural complications were similar in both groups.

CONCLUSIONS

The use of contact force sensing technology is able to significantly reduce ablation, procedure, and fluoroscopy times as well as dose in RFCA of AF in a mixed case-matched group of paroxysmal and persistent AF. Energy delivery is substantially reduced by avoiding radiofrequency ablation in positions with insufficient surface contact. Additionally 12-month outcome data showed increased efficacy. Such time saving and equally safe technology may have a relevant impact on laboratory management and increased cost effectiveness.

Для цитирования: Sigmund Elisabeth, Puererfellner Helmut, Derndorfer Michael, Kollias Georgios, Winter Siegmund, Aichinger Josef, Nesser Hans-Joachim, Martinek Martin. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement-a case-matched comparison in 198 patients.. Pacing and clinical electrophysiology : PACE. 2014. DOI: 10.1111/pace.12549.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.