MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/pace.12157

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2013

Impact of cardiac computed tomography of the interatrial septum before pulmonary vein isolation.

Kobza Richard, Schoenenberger Andreas W, Cuculi Florim, Zuber Michel, Auf Der Maur Christoph, Buhmann Ralf, Resink Therese J, Erne Paul

Контролируемое клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Multidetector computed tomography (MDCT) may be useful to identify patients with patent foramen ovale (PFO). The aim of this study was to analyze whether a MDCT performed before pulmonary vein isolation reliably detects a PFO that may be used for access to the left atrium.

METHODS AND

RESULTS

In 79 consecutive patients, who were referred for catheter ablation of symptomatic paroxysmal or persistent atrial fibrillation (AF), the presence of a PFO was explored by MDCT and transesophageal echocardiography (TEE). TEE was considered as the gold standard, and quality of TEE was good in all patients. In 16 patients (20.3%), MDCT could not be used for analysis because of artifacts, mainly because of AF. On TEE, a PFO was found in 15 (23.8%) of the 63 patients with usable MDCT. MDCT detected six PFO of which four were present on TEE. This corresponded to a sensitivity of 26.7%, a specificity of 95.8%, a negative predictive value of 80.7%, and a positive predictive value of 66.7%. The receiver operating characteristics curve of MDCT for the detection of PFO was 0.613 (95% confidence interval 0.493-0.732).

CONCLUSIONS

MDCT may detect a PFO before pulmonary isolation. However, presence of AF may lead to artifacts on MDCT impeding a meaningful analysis. Furthermore, in this study sensitivity and positive predictive value of MDCT were low and therefore MDCT was not a reliable screening tool for detection of PFO.

Для цитирования: Kobza Richard, Schoenenberger Andreas W, Cuculi Florim, Zuber Michel, Auf Der Maur Christoph, Buhmann Ralf, Resink Therese J, Erne Paul. Impact of cardiac computed tomography of the interatrial septum before pulmonary vein isolation.. Pacing and clinical electrophysiology : PACE. 2013. DOI: 10.1111/pace.12157.