DOI 10.1364/ao.51.005244

Медицинская библиотека · Applied optics · 2012

Numerical analysis of astigmatism correction in gradient refractive index lens based optical coherence tomography catheters.

Wang Tianshi, van der Steen Antonius F W, van Soest Gijs

СтатьяПеревод пока не подготовлен

Аннотация

Endoscopic optical coherence tomography (OCT) catheters comprise a transparent tube to separate the imaging instrument from tissues. This tube acts as a cylindrical lens, introducing astigmatism into the beam. In this report, we quantified this negative effect using optical simulations of OCT catheter devices, and discuss possible compensation strategies. For esophageal imaging, the astigmatism is aggravated by the long working distance. For intracoronary imaging, the beam quality is degraded due to the liquid imaging environment. A nearly circular beam profile can be achieved by a curved focusing optics. We also consider the method of matching refractive indices, and it is shown to successfully restore a round beam.

Для цитирования: Wang Tianshi, van der Steen Antonius F W, van Soest Gijs. Numerical analysis of astigmatism correction in gradient refractive index lens based optical coherence tomography catheters.. Applied optics. 2012. DOI: 10.1364/ao.51.005244.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.