MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/mrm.70326

Медицинская библиотека · Magnetic resonance in medicine · 2026

Low-Cost and Detunable Wireless Resonator Glasses for Enhanced Eye MRI With Concurrent High-Quality Whole-Brain MRI.

Lu Ming, Yang Xiaoyue, Moore Jason E, Li Pingping, Anderson Adam W, Gore John C, Smith Seth A, Yan Xinqiang

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To develop a wearable wireless resonator glasses design that enhances eye MRI signal-to-noise ratio (SNR) without compromising whole-brain image quality at 7 T.

METHODS

The device integrates two detunable LC loop resonators into a lightweight, 3D-printed frame positioned near the eyes. The resonators passively couple to a standard 2Tx/32Rx head coil without hardware modifications. Bench tests assessed tuning, isolation, and detuning performance. B1 + maps were measured in a head/shoulder phantom, and SNR maps were obtained in both phantom and in vivo experiments.

RESULTS

Bench measurements confirmed accurate tuning, strong inter-element isolation, and effective passive detuning. Phantom B1 + mapping showed negligible differences between configurations with and without the resonators. Phantom and in vivo imaging demonstrated up to a ∼3-fold SNR gain in the eye region, with no measurable SNR loss in the brain.

CONCLUSION

The wireless resonator glasses provide a low-cost, easy-to-use solution that improves ocular SNR while preserving whole-brain image quality, enabling both dedicated eye MRI and simultaneous eye-brain imaging at ultrahigh field.

Для цитирования: Lu Ming, Yang Xiaoyue, Moore Jason E, Li Pingping, Anderson Adam W, Gore John C, Smith Seth A, Yan Xinqiang. Low-Cost and Detunable Wireless Resonator Glasses for Enhanced Eye MRI With Concurrent High-Quality Whole-Brain MRI.. Magnetic resonance in medicine. 2026. DOI: 10.1002/mrm.70326.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.