MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jvs.2022.08.016

Медицинская библиотека · Journal of vascular surgery · 2022

One-year safety and effectiveness of the Alto abdominal stent graft in the ELEVATE IDE trial.

Lyden Sean P, Metzger D Christopher, Henao Steve, Noor Sonya, Barleben Andrew, Henretta John P, Kirksey Levester

Многоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

This study reports the results of a prospective, multicenter trial designed to evaluate the safety and effectiveness of the polymer based Endologix Alto Stent Graft System in treating abdominal aortic aneurysms (AAAs), with sealing 7 mm below the top of the fabric in aortic neck diameters from 16 to 30 mm.

METHODS

Seventy-five patients were treated with Alto devices between March 2017 and February 2018 in 16 centers in the United States for infrarenal AAAs (max diameter ≥5.0 cm in diameter or size increase by 0.5 cm in 6 months or diameter ≥1.5 times the adjacent normal aorta). Patients were followed for 30 days, 6 months, and 1 year by clinical evaluation and computed tomography and abdominal x-ray imaging. Treatment success was defined as technical success and freedom from AAA enlargement, migration, type I or III endoleak, AAA rupture or surgical conversion, stent graft stenosis, occlusion, kink, thromboembolic events, and stent fracture attributable to the device requiring secondary intervention through 12 months. Preoperative characteristics, perioperative variables, follow-up clinical evaluations, and radiographic examination results through the first 1 year were analyzed.

RESULTS

The mean patient age was 73 years, with 93% of patients being male. The 30-day major adverse event rate was 5.3%. At 1 year, the primary endpoint was met with a treatment success rate of 96.7%. Through 1-year post-treatment, all-cause mortality was 4.0%. No AAA-related mortality occurred. AAA enlargement was 1.6%, type I endoleak rate was 1.4%, with 100% freedom from type III endoleaks, device migration, device fracture, stent occlusion, or AAA rupture. The device-related secondary intervention rate was 2.7%.

CONCLUSIONS

This prospective study demonstrates the Endologix Alto is safe and effective in treating AAAs with appropriate anatomy at 1 year. The safety endpoint is met by a 5.3% 30-day major adverse event rate, whereas the effectiveness endpoint is met by a treatment success rate of 96%.

Для цитирования: Lyden Sean P, Metzger D Christopher, Henao Steve, Noor Sonya, Barleben Andrew, Henretta John P, Kirksey Levester. One-year safety and effectiveness of the Alto abdominal stent graft in the ELEVATE IDE trial.. Journal of vascular surgery. 2022. DOI: 10.1016/j.jvs.2022.08.016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.