First Case of Nellix Stent Rupture in a Huge Symptomatic Abdominal Aortic Aneurysm Who Underwent Endovascular Aneurysm Sealing 7 Years Before.
Аннотация
PURPOSE: Despite encouraging early results, mid- and long-term follow-up of endovascular aneurysm sealing (EVAS) has shown increased rates of failure mainly associated with endoleak detection and progressive bag separation with aneurysm reperfusion.
CASE REPORT: We present the first case of a Nellix endograft stent fracture detected in a 91-year-old male patient, presenting with widespread abdominal pain, 7 years after elective treatment of an abdominal aortic aneurysm by EVAS. Considering the sudden and unexpected nature of the event, an in-depth analysis of the possible causes of this structural failure has been performed.
CONCLUSION
Material fatigue could be another significant cause of late EVAS failure and should be carefully assessed in addition to endoleak detection during follow-up.
CLINICAL IMPACT: The case presented in this article further underlines the importance of a strict long term follow-up protocol in every patients who underwent EVAS.