Calibration of ipsilateral stimulus transducer for acoustic reflex measurements.

Olsen S, Osterhammel P A, Rasmussen A N, Nielsen L H

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

Pure-tone Reference Equivalent Threshold Sound Pressure Level (RETSPL) of the ipsilateral stimulus receiver for acoustic reflex measurements on Madsen Electronics type Zodiac 901 impedance audiometer is provided. The results, obtained from 20 normal-hearing subjects, are achieved by comparing hearing threshold levels measured using a TDH 39 telephone (calibrated to ISO 389) with thresholds recorded using the ipsilateral stimulus insert phone. The calibration is referenced to an IEC-711 ear simulator and comprises the following frequencies: 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, 8000 Hz.

Для цитирования: Olsen S, Osterhammel P A, Rasmussen A N, Nielsen L H. Calibration of ipsilateral stimulus transducer for acoustic reflex measurements.. Scandinavian audiology. 1995. DOI: 10.3109/01050399509047530.