DOI 10.1167/iovs.18-25473

Медицинская библиотека · Investigative ophthalmology & visual science · 2019

Regional Variations and Intra-/Intersession Repeatability for Scotopic Sensitivity in Normal Controls and Patients With Inherited Retinal Degenerations.

Bennett Lea D, Metz Georgiana, Klein Martin, Locke Kirsten G, Khwaja Areeba, Birch David G

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: Dark-adapted visual fields were obtained from patients with inherited retinal degeneration (IRD) and controls to evaluate the effect that age, retinal region, and disease had on scotopic sensitivity. Intra- and intersession test-retest repeatabilities for patients and controls were measured to establish significant change for longitudinal studies.

METHODS

A total of 41 patients with IRD and 30 controls had one eye dilated and dark-adapted for 40 minutes. Scotopic sensitivity was measured with a Medmont dark-adapted chromatic (DAC) perimeter (size V stimulus, 200-ms duration, background luminance < 0.0001 cd/m2, dynamic range 0-75 decibel [dB]). Mixed effects analysis was performed to analyze age, retinal eccentricity, and sensitivity. The intra-/intersession coefficients of repeatability (CR) were calculated for controls and patients with IRD.

RESULTS

Each additional year was associated with lower sensitivity (-0.22 dB) per year in normal controls over age 50 compared to younger controls (12-49 years). The superior field had lower sensitivity than the inferior, but the nasal field was not different compared to the temporal field in normal controls. The CR for intra- and intersession testing on mean sensitivity (MS)/pointwise sensitivity (PWS) were ±1.5/±8.5 and ±3.3/±9.8 dB, respectively, for patients with IRD. Control MS/PWS CR were ±1.5/±6.1 dB for intrasession and ±1.7/±6.8 dB for intersession DAC perimetry.

CONCLUSIONS

The DAC perimeter is an important asset because it tests a wide field of scotopic vision. The CR are comparable to those of other perimetry devices. Effects of age and retinal region should be considered when assessing scotopic sensitivity measured with the DAC perimeter.

Для цитирования: Bennett Lea D, Metz Georgiana, Klein Martin, Locke Kirsten G, Khwaja Areeba, Birch David G. Regional Variations and Intra-/Intersession Repeatability for Scotopic Sensitivity in Normal Controls and Patients With Inherited Retinal Degenerations.. Investigative ophthalmology & visual science. 2019. DOI: 10.1167/iovs.18-25473.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.